Class II — Moderate risk. May cause temporary or reversible harm; serious harm unlikely.
Why it was recalled
Cross Contamination With Other Products: Oral care solutions were manufactured by a third party supplier on equipment shared with non-pharmaceutical products
What was recalled
Q.Care Petite Size Oral Cleansing & Suctioning System. Contains 1 Covered Yankauer, Suction Handle and Y-Connector, 2 Petit Size Suction Toothbrush packages with Antiseptic Oral Rinse, 4 Petit Size Untreated Suction Oral Swab packages with Antiseptic Oral Rinse, 2 8 French Oropharyngeal Suction Catheter packages. Sage Products, Inc., 3909 Three Oaks Road, Cary, IL 60013. Product Code 6704 UPC 6 18029 84123 0
Recalling firm
Sage Products Inc
Origin
IL, United States
Initiated by
Voluntary: Firm initiated
FDA recall number
D-0393-2018
What should you do? A recall doesn't always mean your specific bottle is affected - it often applies to certain lots only. Check the lot number on your package against the FDA notice, and talk to your pharmacist before stopping any medication.
Reason Cross Contamination with Other Products: due to a potential for cross-contamination with Acemetacin API due to an issue at the API manufacturer.