Curaplex Epi Safe Kit, 8600-01102 TRAINING ONLY, Rx Only, Distributed by Sarnova HC, LLC…
Class II — Moderate risk. May cause temporary or reversible harm; serious harm unlikely.
Why it was recalled
Labeling Incorrect Instructions: This recall has been initiated because the Instructions for Use included in the Epi-Safe Kits for the Epi-Safe Syringe (and/or the Safety Lok Syringe in Model 8600-01120) recommend a midpoint dosage of epinephrine for children between 0 lbs. and 66 lbs. of 0.15mL that has not been approved by FDA for children. Reliance on the current IFU and/or use of the Epi-Safe Syringe as instructed may result in the administration of an inappropriate dosage of epinephrine to young children experiencing anaphylaxis, especially those under 33 lbs. Additionally the Epi Safe syringe does not have the correct markings to allow a health care provider to draw up the correct dosage for children under 33 kg.
What was recalled
Curaplex Epi Safe Kit, 8600-01102 TRAINING ONLY, Rx Only, Distributed by Sarnova HC, LLC s family of companies: Bound Tree Medical, LLC, Cardio Partners, Inc., Emergency Medical Products, Inc. & Tri-anim Health Services, Inc. 5000 Tuttle Crossing Blvd. Dublin, OH 43016
- Recalling firm
- Bound Tree Medical, LLC
- Origin
- TX, United States
- Initiated by
- Voluntary: Firm initiated
- FDA recall number
- D-0385-2019
What should you do? A recall doesn't always mean your specific bottle is affected - it often applies to certain lots only. Check the lot number on your package against the FDA notice, and talk to your pharmacist before stopping any medication.
Get an alert if your medication is recalled
Tell us which medications you take. If the FDA recalls one, we email you - in plain English, with what to do. Free.